Precautions for on-site safe operation of vortex flowmeter

1. Mechanical installation safety matters1. Do not disassemble the instrument under pressure: During maintenance work, if you want to disassemble the instrument, you must check in advance whether there is pressure in the original pipeline. If there is pressure, you must close the upstream valve and downstream valve, and open the pressure relief valve to relieve the pressure, and confirm the pipeline There is no pressure to unload the meter.2. If there is a toxic or corrosive medium in the original pipeline, protective measures must be taken during the disassembly process to prevent poisoning and pollution to personnel and the environment.3. If the pipeline contains flammable and explosive media, protective measures should be taken during installation and disassembly.2. Electrical installation and maintenance safety matters1. When wiring and disconnecting the flowmeter, do not work with electricity.2. Do not open the front and rear cover of the converter when the power is on during maintenance on the explosion-proof site.3. For the vortex flowmeter whose negative terminal of the power supply cannot be grounded, check the insulation resistance between the negative terminal of the power supply and the ”ground” before powering on after wiring, and the insulation resistance is not less than 20MΩ.4. For the vortex flowmeter whose negative end of the power supply cannot be grounded, battery-powered electronic instruments should be used during field testing, and the negative end of the instrument cannot be grounded. The oscilloscope used must have a differential input terminal, and it should be placed in the AC coupling mode when measuring.